3000°C Laboratory Graphitization Furnaces

  • 3000°C Laboratory Graphitization Furnaces
  • 3000°C Laboratory Graphitization Furnaces
3000°C Laboratory Graphitization Furnaces 3000°C Laboratory Graphitization Furnaces

3000°C Laboratory Graphitization Furnaces

Product Introduction:

The graphitization furnace with integrated design is specially designed for laboratories, and is especially suitable for placing in places with high space requirements such as university laboratories. This model can reach an ultra-high processing temperature of 3000°C, and is an ideal choice for the development and research of various new materials.



Applications:

Used in the carbonization and graphitization of carbon materials, such as high-temperature treatment of battery anode materials, high-temperature treatment of graphite powder, carbonization and graphitization of carbon fibers. It is also suitable for high-temperature sintering of precision ceramics, high-temperature components, high-temperature carbonization sintering of powder metallurgy materials, tungsten carbide and other products, and other materials that can be sintered and smelted in a carbon environment.


Features:

● Maximum long-term continuous working temperature: 3000℃ (more than 12 hours)

● Maximum design temperature: 3200℃ (maximum temperature continuous working time <2 hours)

● Working atmosphere in the furnace: nitrogen, argon

● Furnace working chamber size: Φ200×300mm (internal charging space)

● Rated output power: 60KW, the actual power used is not more than 50KW

● Rated intermediate frequency: 4000Hz

● Temperature measurement method: Double color infrared thermometer instrument and equipment

● Temperature measurement range: 1000-3200℃, (0-3200℃ full range temperature measurement, please specify when ordering, the standard configuration starts at 1000℃)

● Temperature control method: automatic program temperature control

Equipment use: high temperature graphitization

● Cooling water consumption: 6~10m3/h

● Cooling water pressure: 0.15~0.25MPa

● Water quality of internal circulating water: distilled water or purified water

● Power supply grid: three-phase five-wire system 380V, 50Hz/60Hz

● Cooling protection: using flow, pressure, water temperature to implement multiple protections for cooling water, etc.

● Over-temperature, over-pressure, gas status, water lack and power failure alarms and automatic mechanical interlock for over-pressure protection.

● Limit heating speed: less than 60 minutes.